Autobrinometer
Autobrinometer
dc.contributor.author | Vijayabharathi, K. | |
dc.contributor.author | Ramakrishnan, K. | |
dc.contributor.author | Sivadas, T.K. | |
dc.date.accessioned | 2013-06-26T08:55:34Z | |
dc.date.available | 2013-06-26T08:55:34Z | |
dc.date.issued | 1982 | |
dc.description.abstract | An instrument developed for the rapid and accurate measurement of brine concentrations during blanching without disturbing the routine blanching operation is described. The concentration is sensed by a plantinum electrode conductivity cell and displayed in a moving coil meter after conversion of the electric signals into D.C. voltage. The instrument can measure in the range 5 to 12% with an accuracy of plus or minus 1%. The errors caused mostly are those due to wide temperature variations of the brine between 95 to 102 degree C and the unknown quantities of protein. | en_US |
dc.identifier.citation | Fishery Technology 1982:19(2), 105-107 | en_US |
dc.identifier.uri | http://hdl.handle.net/123456789/262 | |
dc.language.iso | en | en_US |
dc.publisher | Society of Fisheries Technologists (India) | en_US |
dc.subject | autobrinometer | en_US |
dc.title | Autobrinometer | en_US |
dc.type | Article | en_US |